›› 2019, Vol. 39 ›› Issue (4): 68-.doi: 10.16708/j.cnki.1000-758X.2019.0031

Previous Articles     Next Articles

Study on the inhibition effect of modified Allan variance on Flicker PM noise

 LIU  Ying-Qian, CHEN  Qiang, XUE  Ren-Kui, LAN  Xue-Mei, ZHANG  Zhen-Zhi   

  1. Beijing Orient Institute of Measurement and Test, Beijing 100086, China
  • Received:2018-11-23 Revised:2018-12-24 Published:2019-08-25 Online:2019-04-22

Abstract: In the time and frequency field, Allan variance is utilized in general to indicate frequency stability. Allan variance can not inhibit Flicker PM effectively because it′s defective algorithm. In order to reduce the impact of noise on frequency stability, frequency stability should be indicated by modified Allan variance. Modified Allan variance for inhibition mechanism of Flicker PM is analyzed. Through the satellite commonview method, the comparative data of hydrogen clock and UTC(NIM) are evaluated. Then, the noise model is used to demonstrate the results of evaluation. The effectiveness of the modified Allen variance against Flicker PM inhibition is verified.