Design of Constant Pressure Flowmeter

Zhang Dixin Meng Yang Lü Shiliang (Lanzhou Institute of Physics, Lanzhou 730000)   

  • Published:1999-06-25

Abstract: A constant pressure flowmeter is a kind of calibration apparatus of gas throughput. The calibration range is (17.1~1×10 -5 )PaL/s with the uncertainty less than 2%. It can calibrate the standard leaking hole with the direct method of meassurement and the comparison method of measurement. In the paper, the design method, the structure parameters selection for the constant pressure flowmeter are introduced.